Method Man II
QC You Later
SOP It Like It's Hot
Method to the Madness
M10-Acious
400

Before sample extraction begins, samples should be thawed in this environment

What is an ice bath?

400

This QC is used only during validation

What is the LLOQ QC?

400

This code is used when the wrong expiration date was written and corrected

What is RE (Recording Error)?

400

The minimum number of individual matrix sources evaluated for selectivity

What is six individual, non-pooled sources?

400

This is the primary objective of bioanalytical method validation

What is to demonstrate that the method is suitable for its intended purpose?

800

The acceptable retention time range during system suitability

What is 3.00 to 4.00 minutes?

800

The DQC is prepared from this concentrated solution

What is the 2,000,000 ng/mL stock solution?

800

No pages may be removed from this type of notebook

What is a bound notebook?

800

The maximum allowable IS response in a selectivity blank

What is ≤5% of the average IS response at LLOQ?

800

A chromatographic calibration curve must contain at least this many concentration levels

What is six concentration levels?

1200

The analyte transition monitored for BMS-986278

What is m/z 444.3 → 273.9?

1200

This QC serves as the source solution used to prepare the HQC

What is the 8,000 ng/mL DQC?

1200

If original laboratory records are lost and cannot be reproduced, the resulting study data are considered this

What are unreliable and must be rejected and repeated?

1200

This is what happens if the calibration curve is rejected

What is the entire run must be rejected and reanalyzed?

1200

This is the minimum number of analytical runs required for accuracy and precision validation

What is three runs?

1600

The software used for quantitative data reduction

What is Watson LIMS v7.6.1?

1600

If the DQC were prepared incorrectly, this higher-level QC would also be affected

What is the HQC (800 ng/mL)?

1600

In IDBS, completion of sample-processing steps should generally be documented within this timeframe

What is within 30 minutes of completion of the step?

1600

The minimum number of freeze-thaw cycles that must be evaluated

What is three cycles?

1600

Samples should remain frozen at least this long between freeze-thaw cycles

What is 12 hours?

2000

A 1 mL plate may be used, but this size is preferred

What is a 2 mL 96-well collection plate?

2000

An analyst accidentally used 0.200 mL of stock instead of 0.0200 mL when making the DQC. The DQC concentration would be approximately this much higher

What is 10 times higher than intended?

2000

How we know a notebook page is complete

What is the signed/initialed signature block and lined-out unused space?

2000

Calibration standards and QC samples should normally be prepared from these

What are separate stock solutions?

2000

For chromatographic methods, at least this percentage of ISR results must agree within ±20%

What is two-thirds (2/3)?