Access Management
OR Workflow
Inventory Management
Site Review
SPD
100

This Stakeholder is most important in the implementation of Expanded Access.

Who is the Surgeon Champion?

100

This is how Intuitive's turnover time definition typically differs from the Hospital's

What is da Vinci undock to dock vs Patient Wheels Out to Wheels in Turnover

100

This is the minimum amount of crucial inventory to keep on hand.

What is PAR?

100

These are the electrical requirements for an Xi system

what are3 dedicated circuits (4 if dual consoles) with 15amps each?

100

What are the steam sterilization parameters for the Ion catheter guide?

What is 270F, 4 minutes exposure, and 20-30 minutes dry time. 

200

This is the typical block release time for benign case types such as URO and GYN

What is two weeks?

200

Genesis recommends these parallel tasking activities with the patient in the room.

What is drape da Vinci?

What is setup the backtable?

What is count?

200
This is the number of reprocessing cycles each instrument has

What is five more than the OR lives?

200

These are the low temp sterilizers validated for Xi endoscopes

What are Sterrad 100s, 100nx, Steris VPro Max, VPRO-1, VPRO 1 Plus, VPRO 2
200

These are the hangtime requirements for Ion products

What is it is up to hospital policy, Intuitive does not have a recommendation

300

These are the reasons why Genesis da Vinci utilization calculations are usually lower than that of hospital utilization calculations?

What is OR utilization vs da Vinci utilization?

What is da Vinci utilization does not include 1st case setup to docking time?

300

Genesis recommends the room is ready for the patient when

Equipment checked.

Anesthesia ready.

Bed ready.

Backtable pack opened and tech is scrubbed in.

Trays, Instruments, Disposables available.

300

When calculating par levels, what is the mode and how do you select it? 

The mode is the most number of cases performed in one day that is not a statistical outlier, and is found in the carrying cost tab of excel. 

300

These Five AER's are Validated for Ion Reprocessing

What are Medivator Advantage Plus, Medivator Advantage Pass Through, Medivator DSD Edge, Steris 1E, Steris 1 Endo

300

What is the max temperature allowed for the Xi endoscope? 

What is 140F. 

400

This is the maximum recommended console time and the minimum recommended number of cases needed for flip rooms.

What is less than 90 min and at least 5 cases?

400

Here are three examples of verbal/electronic communication best practices Genesis recommends for the end of the case.

Surgeon communicating with Anesthesia approximate case end time.

RN communicating with EVS approximate wheels out time.

Surgeon communicating with OR Staff his/her target return to room time.

OR Staff communicating with SPD on tray, instrument, or endoscope rapid turnaround requests.

RN communicating with Pre-op next patient status.

400

When calculating par levels, what is the standard selection for turnover time? 

What is 24 hours. 

400

The SP system will fit in most elevators, as long as they are this wide

What is 41in wide?

400

This is the most common reason for disc breakage

What is an enzymatic issue?

500

These are three ways to maximize da Vinci access on a daily basis

What is stack da Vinci cases during their all-day robotic block and stack non-da Vinci cases during other block times.

What is front load robotic cases and putting other procedures at end of the day.

What is movie lap procedures that can be done with da Vinci to da Vinci room during the day when da Vinci becomes available.

What is schedule the same procedure type and room configurations back-to-back, reducing workflow inefficiencies time and the need to rearrange a room multiple times a day.

500

These are four examples of surfaces/equipment that are "enhanced" or "terminal" cleaned, but do not need to be cleaned during each turnover.

What are:

steps, workstation on wheels, garbage or linen bins, stools, walls, doors.

500

For an instrument with 15 lives, what is the minimum utilization percentage to be recommended for trays. 

What is 75% (15 OR lives divided by 20 reprocessing lives). 

500

This is the process ultrasonic's use to clean instrumentation, and how it works.

What is Cavitation, and by sending vibrations through the water, which create bubbles that implode, breaking up debrief?

500

This Amount of Air pressure should be applied when doing a leak test for the SP Camera

What is 140-180mmHg (19-24kpa)?