D. diode.
D. Diode
The atypical (not common) tube mounting design
a. Ceiling mounted
b. Table mounted
C. All of these are common designs
d. Floor mounted
b. Table mounted
With modern radiographic systems, images are now comprised of electronic data sets in a digital format. The data files can be altered and optimized in appearance following exposure and image display. This step is often referred to as
D. postprocessing.
D. Post processing
B. they must have a small electrical charge applied to the TFT layer.
C. wireless detectors use batteries for charging.
D. All of these
D. All of these
In radiation therapy, palliative treatment plans are intended to
B. merge diagnostic studies in a “fusion” type of treatment therapy.
C. eliminate all cancers in the patient’s body.
D. control cancer tissue growth to lengthen and improve the patient’s life.
D. Control cancer tissue growth To lengthen and improve the patients life.
This modality will likely have the highest diagnostic yield for determining pregnancy stages.
A. Routine radiography
B. CT scans
C. All of these have the same diagnostic yield
D. Ultrasound
D. Ultrasound
The X-ray IR system that does not require a reader to produce the image:
A. Ultrasound
B. DR
C. CR
D.All of these require a reader to produce an X-ray image.
B. DR
To improve the flexibility of a radiographic system, engineers have designed equipment so that the X-ray tube support system and detector assembly
B. can move independently of each other.
C.are always connected to the radiologist suite
D.are isotropic.
B. Can move independently of each other
Part B:

A. Gantry
B. Collimator
C. Generator
D. X-ray Tube
B. Collimator
Fluoroscopy of a fractured hip repair done in the surgical theater is typically accomplished with a
B.mobile MRI unit.
C. mini-CT Scanner
D.portable nuclear medicine unit.
A. Portable C-arm system
The option that is favorable for modern X-ray machine tables:
A. None of these are favorable design features.
B. Must be radiopaque to protect the patient from radiation.
C. Sufficiently flat
D. Must have a head rest for technologist's comfort
C. Sufficiently flat
D.variable height table.
B. Upright cassette holder or Bucky
The person on this list who is not permitted to operate this machine:

A. Radiologist
B. Radiologic Technologist
C.Radiologic Technologist Student
D. Secretary
D. Secretary
C. Positive beam limitation
AEC controls this aspect of the radiographic technique selection:
A. The grid vibration frequency
B. Time measured in seconds
C. Time measured in amperes
D. The voltage
B. Time measured in seconds
The choice that is true about radiographic x-ray room tables:
A.Must be of the fixed design.
B. Should be radiolucent
C. Should be radiopaque to protect the patient from exposure
D. Should not be flat
B. Should be radiolucent
The receptor tray is:

1. A
2. B
3. C
4. D
Select the correct option that pertains to mobile radiography
A.Is never necessary in medical imaging departments.
B. Is convenient since it does not require the use of an IR
C. Is performed for situations where the patient is unable to be brought to the department.
D. Is only performed if the patient is too tired to to come to the department
C. Is performed for situations where the patient is unable to be brought to the department
In a radiographic/fluoroscopic (R/F) system, the primary barrier is considered to be the
B. tabletop of the fluoroscopic table.
C. the IR or flat-panel detector.
D. lead apron worn by the fluoroscopist
C. The IR or flat-panel detector
Images acquired in a CT scan are most commonly obtained in the patient’s ____________ plane.
B. sagittal
C. mid-sagittal
D. coronal
A. Axial
The purpose of grid:

A.Absorb scatter that has been produced in the patient during exposures.
B. Reduce the production of scatter within the patient during exposure
C.Eliminate the need for lead shielding.
D.Reduce radiation exposures by ALARA
A. Absorb scatter that has been produced in the during exposures
Physicians can choose from several medical imaging modalities to determine their patient’s condition. Among these are radiography, MRI, CT, ultrasonography, and nuclear medicine, to mention a few. Each specialty has its own _______ of information.
B.diagnostic yield
C.image fidelity
D.interpretation cycle
A diagnostic yield
Radionuclide imaging (nuclear medicine) is very unique in that its primary clinical value is the ability to
B.measure bone mineral density and quantify a FRAX value
D.create high spatial resolution images of bony anatomy that rival CT scans.
C. Assess tissue perfusion and metabolic activities
A key distinctive characteristic of MR scanning is that the source of the MR signal to create the image is the
B. patient.
C. gradient coil.
D.external coil.
B. patient
Desirable features of a modern radiographic table consist of which of the following?
1. High-attenuation tabletop material
2. Variable height
3. Four-way floating tabletop
4. Radiolucent tabletop
5. Convenient and ergonomic-friendly controls
6. Automatic filtering of high energy photons
B. 2, 3, 4, and 5 only
D. 1, 2, 3, 5, and 6 only
B. 2, 3, 4,and 5 only