Quintet AC Glucometer
Capillary Fingerstick Collection
Siemens DCA A1C
UriScan Optima
Abbott Covid/Hemocue Hb
100
Temperature storage for Control Solutions and Test Strips.

What is Room Temperature?

100

Patient's hand status for best blood flow.

What is patient's hands are warm, relaxed and no rings?

100

Reported Hemoglobin value reported in EPIC.

What is in percentage?

100

Report in Epic when Urobilinogen is "norm".

What is 0.2?

100

Stability of Opened Eurotrol Quality Controls.

When is 30 days and refrigerated?

200

Symbol in need to run Quality Control in Glucometer.

What is "CS" symbol?

200

Drops to wipe away. 

What are 2-3 drops?

200

Expiration for reagent kits stored at room temperature.

What is 3 months or the expiration date printed on the box, whichever comes first?

200

Daily Maintenance actions.

What is Clean the Strip Holder and Document on the UriScan maintenance log?

200

Daily duties of Hemocue Hb 201+ Analyzer.

What is Clean Cuvette Holder and document on the Maintenance Log?

300

Expiration date on Opened Test Strips.

When in 3 months from the open date or the expiration date on the bottle, whichever one comes first?

300

Fingers to avoid.

What is the 5th, index and thumb?

300

Calibration and Quality Control frequency.

When is new lot and new shipment of reagents?

300

Acceptable time from collection to analyze urine with the UriScan. 

What is 1 hour?

300

Acceptable Specimen for Abbott ID Covid-19.

What is nasal swab?

400

Expiration date on Opened Blood Glucose Controls.

When in 3 months from the open date or the expiration date on the bottle, whichever one comes first?

400

Ideally used hand for collection.

What is Non-Dominant Hand?

400

Stability for Reconstituted Controls.

What are 3 months?

400

Stability for Opened BioSys Liquid Urine Controls.

What is 20 dips?

400

Frequency for Quality Control on Abbott ID Covid Instrument.

When is New Lot Number, New Shipment, Instrument Relocation and in the event of Power Outage?

500

Glucometer Control Reference Ranges location (find at). 

Where in the Test Strip Canister?

500

Fingerstick sample for best blood flow and least pain. 

Where is at the side of the finger?

500

The Weekly Duties for Siemens A1C Instrument.

What is Clean the Onboard Barcode Reader Window and Exterior?

500

Uriscan Quality Control Frequency.

When is new lot of test strips, opening new test strips and if canister is in use longer than 30 days?

500
Daily Maintenance actions for Abbott Covid ID Instrument.

What is clean the exterior of the instrument, the surface visible under the open lid and Clean surrounding bench area before and after use?