What is Room Temperature?
Patient's hand status for best blood flow.
What is patient's hands are warm, relaxed and no rings?
Reported Hemoglobin value reported in EPIC.
What is in percentage?
Report in Epic when Urobilinogen is "norm".
What is 0.2?
Stability of Opened Eurotrol Quality Controls.
When is 30 days and refrigerated?
Symbol in need to run Quality Control in Glucometer.
What is "CS" symbol?
Drops to wipe away.
What are 2-3 drops?
Expiration for reagent kits stored at room temperature.
What is 3 months or the expiration date printed on the box, whichever comes first?
Daily Maintenance actions.
What is Clean the Strip Holder and Document on the UriScan maintenance log?
Daily duties of Hemocue Hb 201+ Analyzer.
What is Clean Cuvette Holder and document on the Maintenance Log?
Expiration date on Opened Test Strips.
When in 3 months from the open date or the expiration date on the bottle, whichever one comes first?
Fingers to avoid.
What is the 5th, index and thumb?
Calibration and Quality Control frequency.
When is new lot and new shipment of reagents?
Acceptable time from collection to analyze urine with the UriScan.
What is 1 hour?
Acceptable Specimen for Abbott ID Covid-19.
What is nasal swab?
Expiration date on Opened Blood Glucose Controls.
When in 3 months from the open date or the expiration date on the bottle, whichever one comes first?
Ideally used hand for collection.
What is Non-Dominant Hand?
Stability for Reconstituted Controls.
What are 3 months?
Stability for Opened BioSys Liquid Urine Controls.
What is 20 dips?
Frequency for Quality Control on Abbott ID Covid Instrument.
When is New Lot Number, New Shipment, Instrument Relocation and in the event of Power Outage?
Glucometer Control Reference Ranges location (find at).
Where in the Test Strip Canister?
Fingerstick sample for best blood flow and least pain.
Where is at the side of the finger?
The Weekly Duties for Siemens A1C Instrument.
What is Clean the Onboard Barcode Reader Window and Exterior?
Uriscan Quality Control Frequency.
When is new lot of test strips, opening new test strips and if canister is in use longer than 30 days?
What is clean the exterior of the instrument, the surface visible under the open lid and Clean surrounding bench area before and after use?