steps to record OAEs
DPOAE protocols
DP gram key
Noise Floor
decrease noise
100

step 1-3

1. otoscopy, instruct listener, choose protocol, place probe tip with seal, test in quiet room

2. choose test type: DP or TE

3. enter demographic data 

100

general diagnostic frequencies

500-8000 Hz

100

X axis

f2 frequency in Hz 

100

noise floor may show

95th percentile noise level for normal adults

100

physiologic noise

blood flow, respiration, body movements, chewing, talking, crying

200

step 4 and 5 

monitor stimulus levels

monitor noise levels

200

infant hearing screening frequencies

2000-4000 Hz

200

Y axis

amplitude of OAE response in dB 

200

where do we want noise floor

below 95th percentile

NF amplitude should be at or below -10 dB SPL 

200

ambient noise

room from air condition, tester noise

300

step 6

replicate OAE measurement in same ear

300

L1 is

65 dB

300

red circles 

green triangles

blue lines

gray lines

red circles: OAE response amplitude as function of f2; reflects cochlear health on basilar membrane corresponding to f2

green triangles: noise floor

blue lines: normal OAE amplitude range

gray lines: normal noise floor range

300

equipment noise

decrease by? 

electrical (lights) 

decrease by turning them off

400

step 7-8

change to other ear

replicate OAE measurement in other ear

400

L2 is 

55 dB

400

top 95th percentile

if OAE exceeds this for those with loss, then there is less than a 5% chance the OAE came from ear with loss

400

decrease physiologic noise

instruct to remain still, provide distractors for young kids, children asleep, stabilize cord to prevent movement of probe, increase amount of signal averaging during recording 

500

step 9

change protocol as indicated

500

frequency ratio

1.2

500

bottom line is 5th percentile

if OAE is below this with normal hearing, there is less than a 5% chance it came from an ear with normal hearing 

500

decrease ambient noise

close test room door, position pt far from equipment, may be possible to locate pt in booth and OAE equipment outside the booth, turn pt so test ear is away from equipment, re-seal OAE probe-tip deeper in ear canal, increase amount of signal averaging during OAE recording