This liquid is commonly confused with a blood spill
What is the PCS solution
The first step before troubleshooting a GEM 5000
What is download the Copy IL Data
This is your one stop for all your needs on the Sigma
What is the Rotem Sigma cartridge
This opens the fluidic paths for the cartridge
What is the Pak Opener
This port on the back of the GEM 5000 will get you moving
What is Port A, RS232
This phrase is the appropriate response when a customer has a GEM cartridge fail
What is "Let's try another cartridge"
Severity level 8 refers to what for the Event log
What is not Service Action needed
This is a critical step when replacing the Sample Handler pcb
What is the ADM alignment procedure
These four parts should always be inspected on a GEM 5000 before any hardware troubleshooting
What are the "optic head/heater block mounts"
You are connected to the ETC when
What is "The front display lights up"
This component prevents a GEM cartridge from being used on another GEM
What is the E-prom board
In Service mode, this pump is adjusted first
What is the EC pump
This points to what component
What is the rear BD
This pump is on the left
What is the CoOx pump
This is important to have in the GEM when completing the CoOx Calibration
What is "Nothing"
PVP should be stored at this temperature before using
What is Room Temp
This component is not in the back of the GEM
What is the Rear Fan
You are =+/- 300
What is "center value for the ADM"
What is the Rotary Valve
This tool has the same name as a popular Disney Movie
What is the "Pinocchio Gauge"
This important sensor is located in the GEM cartridge
What is the probe sensor
Never run this test first
What is Test 2
This is used to verify the Sigma weekly
What is the System QC Cartridge
This part is a common cause of intra-spect errors
What is the Diverter Valve
This component is removed from the ETC for testing your pumps
Interface Card