Device Basics
Monitoring/Vitals
Cardiac Interventions
Troubleshooting
CPR and Data Documentation
100

How often does Stryker recommend users perform a shift check? 

Daily

100

How many full waveforms can be viewed on screen at once? 

5

100

What is the maximum amount of energy you can deliver on the LP35? 

360J

100

You notice a flat line or dashed line in the ECG channel, what is the first thing you check?

Lead connection/Which lead is selected to be in ECG channel 

100

Which channel/vital can be used as real time physiological feedback during compressions? 

EtCO2 

200

How much monitoring time will one fully charged battery give you? 

About 4.5 hours 

200

True or False: You shouldn't have the BP cuff on a patient's arm when powering the device on.

True

200

What mode does the device pace in as a default? 

Demand Mode

200

Your printer isn't printing. What are two things to check?

1. Make sure there is paper

2. Paper is loaded correctly (beard not mullet)

3. 1 inch of paper out when the printer is closed 

4. Printer cable has a secure connection

200

What is the purpose of CPRinsight? 

Minimize pauses and improve compression fraction

300

True or False: The device cannot be used on a patient if the readiness indicator is blinking. 

False 

300

If you have a 4 lead ECG cable and pads on a patient, what will the device default the ECG channel to read through? 

Pads

300

True or False: The provider will see the underlying rhythm in the ECG channel when CPRinsight is turned on. 

False

300

"SpO2 Sensor Off Patient" message appears on the screen, what are the 3 ways to address this?

1. Put sensor back on patient

2. Reset channel by unplugging and replugging

3. Remove trunk cable and then remove channel from the screen

300

When events are recorded in the LP35, they capture a rhythm strip that is how long? 

8 seconds (3 pre and 5 post)

400

Name one reason why the readiness indicator may be blinking

1. One or more batteries are at or below the Low battery level

2. Unrecognized battery of battery fault detected

3. No batteries or only one battery installed

4. Auto test failed

5. User test failed (Shift check or Monthly check)

6. Therapy cable or internal paddles not connected during auto test or shutdown

7. Test load connected 

400

Which channel will show you the patient's respiratory rate? 

EtCO2

400

True or False: You can cardiovert with only pads on the patient.

True

400

True or False: You can trend vitals while therapies are ongoing. 

False

400

How many times will the device attempt to transmit before officially failing?

3 times 

500

Where on the device is the EtCO2 exhaust port located? 

On the back of the device in the right lower corner 
500

If you are monitoring a pediatric or neonate patient, which setting should you change on the monitor to automatically adjust alarm limits and initial cuff pressure and where is it located? 

Patient type setting under "Alarms" 

500

What happens when a provider hits "pause" during pacing? 

The rate drops to 25% of the set rate to reveal the patients own rhythm. 

500

My readiness indicator is blinking...where on the device do I go to see the exact reason why?

Under the "System" menu, which you can get to by clicking on the battery icons in the header

500

How long does the device store patient cases in the archives?

30 days