Put the “Q” *in* quality imaging
Numeracy
Potpourri
Components
All Acronyms
100

PET radioactivity is given in these SI units

Becquerel

100

This is the HU value of water.

0 (this is by definition/calibration)

100

To reduce heating of the anode, one of these parameters can be reduced

kV, mA, and/or s
100

These are the major components, in order, in an X-ray tube

System is in a vacuum and surrounded by oil to help dissipate heat.  Heated cathode filament emits electrons. Electric field accelerates them to anode (strength of field governs the kVp of the photons). Tungsten (high Z, high melting point) anode spins to dissipate heat from electron beam while ~1% of electrons produce bremsstrahlung photons (and a few characteristic X-rays). X-rays exit through a barium window and go through a filter to remove remaining lower energy photons.

100

FOV and ROI

Field of View and Region of Interest

200

If the mAs is doubled, this is how much the SNR in an x-ray image improves

Square root of 2

200

This is the typical range of kVp for clinical diagnostic CT scanners

80 – 140 kVp

200
Mammograms have better resolution than general x-rays because of these two properties of the X-ray tube

Smaller anode angle and smaller filament (focal spot size f = F sin(θ))

200

These are the major components, in order, in PET imaging

Radioactive drug (injected, ingested, inhaled), patient, detector crystals (e.g., LYSO), Photomultiplier tube (PMT), electronics, reconstruction computer

200

ADC

Analog to digital converter

300

This term describes the set of RF pulses and gradient fields used to generate an MRI image

Sequence

300

42.58 MHz/T is what quantity?

Gyromagnetic ratio of a proton

300

Mean over standard deviation within an ROI is one way how we calculate this image quality metric

Signal to noise ratio (SNR)

300

These are the major components in Computed Tomography

X-ray tube (has higher heat capacity than planar x-ray), filter, collimator (usually limited to 2-3 options), patient on moving bed, anti-scatter grid, detectors (thin and curved to match curve of bore), reconstruction computer

300

AEC

Automatic Exposure Control

400

Multiplying the magnetic field B0 by the gyromagnetic ratio provides this quantity used in MRI

Larmor Frequency

400

An image with this many grayscale levels has a bit depth of 8

256

400

Ghosting, distortion, motion and aliasing are artifacts in what imaging modality?

MRI

400

These are the major components in Ultrasound imaging

Transducer with piezoelectric crystal, coupling gel, patient, coupling gel, transducer, processing electronics and monitor
(signals are sent out from the transducer, echo back and are detected by the same transducer)

400

DR and CR and TFT

Digital Radiography (directly digitized, read out with Thin-Film Transistor) and Computed Radiography (detector cassette + reader to digitize)

500

Tube arcing and ring artifacts are this type/category of CT artifact

Equipment

500

This is the rest mass energy of an electron or positron

511 keV (this is thus the energy of each of the resulting photons used in PET from positron-electron annihilation)

500

These are the two main reasons for collimation in x-ray imaging

Reduce dose and scatter (scatter reduces contrast)

500

These are the major components, in order, in MR imaging

B0 field (from main solenoid electromagnet), patient (protons align on average with the B0 field), RF pulse to excite a single slice (selected with gradient field and matched frequency), phase altering gradient field (orthogonal), frequency altering gradient field (orthogonal), coil receives RF signal

500

CTDI and DLP

Computed Tomography Dose Index (usually weighted based on central and peripheral measurements, adjusted for pitch in helical scans) and Dose Length Product (CTDIvol * Length), proportional to effective dose - need a k-factor to *estimate* patient dose