Suction
Components
Docking
Display Screen
Troubleshooting
100

The Neptune is a high-suction device and can reach up to ____ mmHg

520 mmHg

100

This is located on the back of the Neptune and must be pressed before the system can be used

Power switch

100

This wash cycle takes just 2 minutes 

Quick Wash 

100

This small button controls the IV pole 

Up, down arrows
100

This silly mistake is the most common reason for a Neptune not releasing from the docking station 

The break is locked

200

At best, WALL suction is typically around ____ mmHg

300 mmHg

200

This number is located below the power switch and must be submitted with service requests

Serial number

200

This blue soap must be attached to the docking station and lasts ~ 200 wash cycles before needing refilled 

Docking detergent 

200

This button is green and must be pressed to activate suction

Start suction
200

If you experience weak suction with the Neptune, you should first check this

Manifold ports 

300

With two manifolds inserted, the Neptune can accommodate up to ____ lines of suction

8 LINES

300

This single-patient use disposable must be inserted into the canister before suction can be activated 

The manifold 

300

The "Extended Wash" cycle takes 45 minutes and should be completed every ____ weeks 

2 Weeks

300

This button is red and must be pressed to stop suction

Stop suction

300

The manifold won't activate suction

It is expired (36 hours has elapsed)

400

This suction dial controls the 4L canister

TOP DIAL 

400
This canister is most accurate (+/- 50ml) and has a light 

4L canister

400

These silver prongs at the bottom of Neptune are what latch onto the docking station 

Magnetic couplers

400

This button allows you to "zero out" volumes between patients

Reset Volumes

400

The Neptune is giving a "docking recommendation" error.  It likely has not been docked in ____ hours 

38 hours

500

If you empty the 4L canister into the 20L canister, the system will lose suction for up to ______ seconds

30 seconds 

500

The "HEPA" filter is the internal vacuum filter.  It must be changed after 500 hours of suction, or every ___ months  

6 Months

500

Neptune docking is "hands-free".  If you place the Neptune on the docking station and walk away, it will automatically complete this type of wash cycle

Normal Wash Cycle (5 minutes)

500

This button can drop the contents of the 4L canister into the 20L canister

Empty 4L Canister 

500

The HEPA filter continues to show as expired even though it was changed out. This button must be pressed

Settings > accessories > reset filter hours