The Neptune is a high-suction device and can reach up to ____ mmHg
520 mmHg
This is located on the back of the Neptune and must be pressed before the system can be used
Power switch
This wash cycle takes just 2 minutes
Quick Wash
This small button controls the IV pole
This silly mistake is the most common reason for a Neptune not releasing from the docking station
The break is locked
At best, WALL suction is typically around ____ mmHg
300 mmHg
This number is located below the power switch and must be submitted with service requests
Serial number
This blue soap must be attached to the docking station and lasts ~ 200 wash cycles before needing refilled
Docking detergent
This button is green and must be pressed to activate suction
If you experience weak suction with the Neptune, you should first check this
Manifold ports
With two manifolds inserted, the Neptune can accommodate up to ____ lines of suction
8 LINES
This single-patient use disposable must be inserted into the canister before suction can be activated
The manifold
The "Extended Wash" cycle takes 45 minutes and should be completed every ____ weeks
2 Weeks
This button is red and must be pressed to stop suction
Stop suction
The manifold won't activate suction
It is expired (36 hours has elapsed)
This suction dial controls the 4L canister
TOP DIAL
4L canister
These silver prongs at the bottom of Neptune are what latch onto the docking station
Magnetic couplers
This button allows you to "zero out" volumes between patients
Reset Volumes
The Neptune is giving a "docking recommendation" error. It likely has not been docked in ____ hours
38 hours
If you empty the 4L canister into the 20L canister, the system will lose suction for up to ______ seconds
30 seconds
The "HEPA" filter is the internal vacuum filter. It must be changed after 500 hours of suction, or every ___ months
6 Months
Neptune docking is "hands-free". If you place the Neptune on the docking station and walk away, it will automatically complete this type of wash cycle
Normal Wash Cycle (5 minutes)
This button can drop the contents of the 4L canister into the 20L canister
Empty 4L Canister
The HEPA filter continues to show as expired even though it was changed out. This button must be pressed
Settings > accessories > reset filter hours