Code or Standard?
TJC
NFPA
Miscellaneous
Recent Changes & Emerging Tech
100

This is a voluntary set of technical definitions and guidelines

Standard

100

Standard EC.02.04.01: The hospital manages _______ __________ risks.

medical equipment

100

Per NFPA 99, it is defined as a space, within a location intended for the examination and treatment of patients, extending 1.8 m (6 ft) beyond the normal location of the patient bed, chair, table, treadmill, or other device that supports the patient during examination and treatment and extending vertically to 2.3 m (7 ft, 6 in) above the floor. 

What is Patient Care Vicinity

100

TJC HAP EC.02.04.01 EP4

The hospital identifies the activities and associated frequencies, ___ ________, for maintaining, inspecting, and testing all medical equipment on the inventory. These activities and associated frequencies are in accordance with manufacturers’ recommendations or with strategies on an AEM program.

 in writing

100

Effective January 1, 2022, TJC updated the definition of this key term as follows: 

"A patient safety event (not primarily related to the natural course of the [patient’s] illness or underlying condition) that reaches a [patient] and results in death, severe harm (regardless of duration of harm), or permanent harm (regardless of severity of harm)."

What is a Sentinel Event


200

Staff competence is initially assessed and documented as part of orientation.

standard

TJC HR.01.06.01 EP5

200

The Joint Commission standard requires the hospital to evaluate its medical equipment management plan every __________ months.

12 (annually)

200

This code is also referred to as the National Electrical Code (NEC)

NFPA 70

200

These are the four major CMS approved accrediting organizations for hospitals

The Joint Commission, Det Norske Veritas (DNV), the Center for Improvement in Healthcare Quality (CIHQ), and the Healthcare Facilities Accreditation Program (HFAP)

200

A relocatable power tap (RPT) must not be connected to these

Other RPTs or extension cords

NFPA 99 10.2.3.6(4)

300

User facilities must report a suspected medical device-related death to both the FDA and the manufacturer. User facilities must report a medical device-related serious injury to the manufacturer, or to the FDA if the medical device manufacturer is unknown.

Code 

FDA: Medical Device Reporting (MDR) regulation (21 CFR Part 803) & Safe Medical Devices Act of 1990 (SMDA)

300

This tool was developed to provide health care organizations with the information they need to prioritize resources and focus corrective action plans in areas that are most in need of compliance activities and interventions.

What is the "Survey Analysis for Evaluating Risk” (SAFER) matrix? 


300

For Relocatable Power Taps (RPTs) the sum of the ampacity of all appliances connected to the outlets must not exceed this percent of the ampacity of the flexible cord supplying the RPT

75%

NFPA 99 10.2.3.6(2)

300

These device types are not eligible for AEM

-Medical laser devices

-Imaging and radiology equipment (whether used for diagnostic or therapeutic purposes)

-New medical equipment with insufficient maintenance history

300

This authority allowed the FDA to approve in vitro diagnostics for the detection and/or diagnosis of COVID-19 (February 4, 2020), personal respiratory protective devices (March 2, 2020), and other medical devices, including alternative products used as medical devices (March 24, 2020), for use during the COVID-19 outbreak.  

Emergency Use Authorization (EUA)

Secretary of the Department of Health and Human Services (HHS) determined, pursuant to section 564 of the Federal Food, Drug and Cosmetic (FD&C) Act  

400

Equipment which may become contaminated with blood or other potentially infectious material (OPIM) shall be decontaminated before servicing or shipping unless decontamination is infeasible. If so, a readily observable label in accordance with (g)(1)(i)(H) shall be attached to the equipment and also state which portions of the equipment remain contaminated.

Code 

Occupational Safety and Health Administration’s (OSHA) Bloodborne Pathogens (BBP) standard, 29 CFR 1910.1030.

400

This TJC EP was eliminated in 2017 because it addresses a “routine part of operations of clinical-care processes,” and “organizations already have multidisciplinary teams in place to give input on equipment.”

EC.02.04.01, EP 1 required hospitals to solicit input “from individuals who operate and service equipment when it selects and acquires medical equipment.”

400

According to NFPA 70 (517.20), receptacles within the area of this type of location must have ground-fault circuit-interrupter protection for personnel if interruption of power under fault conditions can be tolerated or be served by an isolated power system if such interruption cannot be tolerated.

Wet procedure location

400

The July 2022 edition of TJC Environment of Care chapter requires that hospitals that use TJC for deemed status purposes must comply with this year edition of NFPA 99

2012 edition of NFPA 99

400

Patient Matched Medical Devices (PMMD) (ex. Ortho Maxio Facial Cutting Guide) that are created using 3D printing/additive manufacturing must follow these standards regardless of the location of their manufacture

FDA Good Manufacturing Practices (GMP) 

500

Following installation of a water treatment, storage and distribution system, the user is responsible for regular surveillance of the microbiology of the system and for complying with the requirements of this document, including those requirements related to action levels.

Standard

ANSI/AAMI/ISO 23500-1:2019

Preparation and quality management of fluids for haemodialysis and related therapies

500

The hospital’s plan for managing its resources and assets describes in writing the actions the hospital will take to sustain the needs of the hospital for up to ___ hours based on calculations of current resource consumptions.

96 hours

TJC EM.12.02.09 EP3


500

For appliances that are used in the patient care vicinity, the resistance between the appliance chassis, or any exposed conductive surface of the appliance, and the ground pin of the attachment plug shall be less than 0.50 ohm under these conditions

When the cord is flexed at its connection to the attachment plug or connector AND its connection to the strain relief on the chassis

NFPA 99 10.3.2.1

500

Prior to a new wireless medical telemetry system being installed in a hospital the system must be registered with this organization to assure frequency coordination

American Society for Health Care Engineering (ASHE) per the Federal Communications Commission (FCC)

500

This is the term FDA uses to categorize medical Artificial Intelligence (AI)/Machine Learning (ML) software

Software as a Medical Device (SaMD)