SHALL
LABEL ME
INSTALLATION
WARNING!!
EASY PEASY
100

SYSTEMS SHALL BE DEEMED THIS AT THE POINT OF PIPELINE INTRUSION.

What is BREACHED

100

PIPING SHALL BE LABELED BY THESE.

What is STENCILING OR ADHESIVE MARKERS.

100

THIS ASTM MARKING SHALL BE ON EVERY LENGTH

OF PIPE FOR POSITIVE PRESSURE MEDICAL GAS SYSTEMS.

What is ASTM B 819.

100

ITS PERMITTED TO HAVE A SINGLE ALARM PANEL IN THESE TWO CATEGORIES.

What is a CATEGORY 2 & CATEGORY 3

100

CUT ENDS SHALL BE.

What is ROLLED SMOOTH OR DE-BURRED .

200

DURING THE STANDING PRESSURE TEST THE SOURCE VALVE SHALL BE THIS.

What is CLOSED.

200

THE PIPING LABEL SHALL INCLUDE THESE THREE THINGS. HINT: ONE IS IF SOMETHING IS NON-STANDARD.

What is the

 NAME OR CHEMICAL SYMBOL.

GAS OR VACUUM SYSTEM COLOR CODE

WHERE POSITIVE PRESSURE IS ABOVE NORMAL OPERATING PRESSURE; THE OPERATING PRESSURE IN ADDITION TO THE NAME OF THE GAS.

200

FIELD  INSTALLED MEDICAL-SURGICAL VACUUM AND WAGD. SHALL BE THESE THREE COPPER TUBES.

What is 

 ASTM B  88

 ASTM B 280

 ASTM B 819

200

IN THIS AREA THE ALARMS SENSORS SHALL BE LOCATED ON THE PATIENT OR USE SIDE OF THE ZONE VALVE BOX ASSEMBLY.

What is CRITICAL CARE AREA.

200

WHILE BRAZING, JOINTS SHALL BE CONTINUOUSLY PURGED WITH THIS GAS.

What is NITROGEN NF.

300

THE MEDICAL AIR INTAKE SHALL BE LOCATED THIS DISTANCE OFF THE GROUND.

What is 20 FT.

300

AN OXYGEN LABEL WOULD BE THIS COLOR.

What is GREEN/WHITE OR WHITE/GREEN

(BACKGROUND/TEXT)

300

FIELD INSTALLED MEDICAL-SURGICAL VACUUM AND WAGD SYSTEMS, IF NOT COPPER, CAN BE THIS MATERIAL.

What is STAINLESS STEEL TUBE.

300

A HIGH MEDICAL AIR DEW POINT WILL INITIATE A SIGNAL AT THESE TWO ALARM PANELS.

What is LOCAL AND MASTER.

300

BRAZING ALLOY SHALL EXHIBIT A MELTING TEMPERATURE OF HOW MANY __________*F

What is 1000*F

400

THE MEDICAL AIR INTAKE SHALL BE LOCATED THIS FAR FROM ANY "NOXIOUS FUMES"

What is 25 FT.

400

A MEDICAL AIR COLOR LABEL WOULD BE THIS COLOR

What is YELLOW/BLACK

(BACKGROUND/TEXT)

400

IF JOINED TO THE VACUUM PIPING, WAGD SYSTEM PIPING SHALL BE CONNECTED AT A MINIMUM DISTANCE FROM ANY VACUUM INLET.



What is 5 FT.

400

ALARM SWITCHES AND SENSORS ARE TO BE INSTALLED TO BE ___________.

What is REMOVABLE.

400

PIPE LABELS SHALL BE PLACED A MINIMUM DISTANCE APART.

What is 20 FT.

500

THIS SHALL SEPARATE THE ZONE VALVE BOX ASSEMBLY AND THE OUTLET/INLET THAT IT SERVES.

What is a WALL

500
A MEDICAL-SURGICAL VACUUM WOULD BE THIS COLOR.

What is WHITE/BLACK

(BACKGROUND/TEXT)

500

ANY POSITIVE PRESSURE MEDICAL GAS SYSTEM OVER 185 PSI SHALL BE ASTM B 819 TYPE ____.

What is K.

5.1.10.1.4*

500

CANCELABLE AUDIBLE INDICATION OF EACH ALARM CONDITIONS SHALL BE A MINIMUM LEVEL OF ____dBA  AT _______FT.

What is 80 dBA at 3 FT.

500

A VALVE THAT RISES ADJACENT TO THE PIPELINE.

What is a RISER VALVE.