Medical Gas Cylinders
Gas Supply Systems
Anesthesia Machine
Vaporizers, CGO, Manifolds, Flowmeters
Breathing Circuits
100

These are the most common gas cylinders used in anesthesia

What is E cylinders and H cylinders

100

These are the three main components of the medical gas pipeline system

What is the 

Central supply

Pipelines to points of use

Connections to equipment

100

These are 3 safety features that prevent hypoxic gas mixtures.

What is the 

Oxygen supply pressure failure alarm

Pressure sensor shut-off (fail-safe)

Gas flow proportioning

100

This mechanical flowmeter is always placed to the right from the other gases for this reason

Mechanical O2 flowmeter is placed downstream to prevent other gases from streaming in if the system is no longer intact.

100

These are 3 of the 6 desirable characteristics of a breathing circuit

What are

Low resistance to gas flow

Minimal rebreathing of exhaled gases

Removal of CO2

Rapid changes of delivery

Warm and humidification

Safe disposal of waste gases

200

Nitrous oxide will read at this PSIG until the liquid runs out. 

What is 745 PSIG?

200

Bulk storage of oxygen for a facility typically falls into three main categories for these types of facilities.

What is 

Liquid oxygen for large facilities

Series of H tanks for small, rural facilities

Oxygen concentrators for deployment or humanitarian missions

200

The low pressure gas system is comprised of the following parts

What are all parts downstream of the gas flow control valves?

200

These are 3 of the 5 aspects of the vaporizer manifold

What are

Located downstream from flowmeters

Removed or added by user

Interlock system to prevent more than 1 vaporizer being used

Prevents carrier gas from entering a non-utilized vaporizer

Low pressure check = manifold check


200

Barolyme or soda lime is part of this component of the anesthesia breathing circuit

What is the CO2 absorber

Color indicator of violet when used up

May regenerate

Irregular surface maximizes surface area

Neutralization results in H2O and heat

300

The DOT requires gas cylinders to have these 3 markings and a tag that indicates what

What is the

Service pressure

Individualized codes

Inspection Date 

Tag (Full, In Use, Empty)

300

These are two noninterchangeable interfaces that are specific to a gas outlet 

What is quick connector (which is not standardized) among manufacturers

DISS (Diameter Index Safety System)

300

This apparatus converts a variable high input gas pressure to a constant, lower output pressure

And when triggered it causes 3 events to occur

What is the pressure regulator?

What are the pressure relief valve

Emitting a loud sound

Loss of O2 pressure

Failure of O2 delivery

300

The common gas outlet is designed not to do the following

What is not be interchangeable with the scavenging system

It is designed to have all gases exit through it; to be used with jet ventilation and have a female fitting


300

This closed breathing circuit is the most common circuit used and it has three required components

What is the circle breathing circuit

What is must have 

FGF

Pop-off valve

CO2 absorber

Advantage = warmth and humidification


400

Cylinder pressure relief valves come in these three types.

What is 

Fusible plug (metal alloy melts)

Frangible disk assembly (breakable metal disk)

Safety relief valve (spring loaded)

400

These 3 apparatuses are safety features that are used to protect the integrity of the medical gas delivery systems.

What are the 

Emergency shut off valve

DISS and quick connectors

Non-compressible hoses

Regulated by NFPA

400

This valve opens directly from the oxygen pipeline or first stage regulator to the common gas outlet at the following flow rates

Oxygen flush valve with flow rates between 35-75 L/min

400

The main electrical switch on the anesthesia will allow what to occur even during a power failure

What is allow gas flow to continue

It also has multiple outlets for anesthesia monitors

But can overload.  

Never let OR staff plug in non-emergency equipment into the anesthesia machine

400

The commonalities of theses circuits are








What is 

No CO2 absorption

No separate inspiratory or expiratory limbs

FGF and APL valve varies in location

500

This is a safety system that matches the yoke to the valve port.

And it has this many different pin positions.

What is the pin index system?

What is 7?

500

These are 4 of 7 hazards that may occur with the medical gas delivery system

What are 

Insufficient pressure

Crossed pipelines

Debris

Depletion of central supply 

Failure of alarms

Pipeline leaks

Frozen regulators

500

The purpose of the oxygen flush valve 

What is

Filling the breathing system by by-passing the vaporizers

Dilutes the anesthetic

Risk of barotrauma

Recessed button

500

What is this image?





Variable bypass vaporizer

Splits FGF between the bypass and the vaporizer

Higher FGF more volatile agent is used

500

This Mapleson D system has a coaxial modification and has a high risk for what.

What is the Bain system and it has a high risk for kinking of the inner tubing and disconnect