What is the COMPLETE proper attire while cleaning instruments?
Face shield, gown, gloves, shoe cover, mask, head cover.
How long does reps should deliver loaner sets if they aren’t emergent?
48 hours prior to the START of the case.
Name of indicator used to identify protein.
Resi-Test
Sets temperature prior storage?
75 degrees
How many sets in a load need to have moisture inside for it to be considered a wet load?
Two or more sets.
If a clinic drop off instruments in a biohazard bin and you notice that point of use treatment hasn’t been started, what is the next step?
Return the biohazard bin to clinic personnel.
Name three things you should do when receiving a loaner sets.
-weighting
-initial inventory
-upload in censitrac
-take photo and upload into system
- make tag for set
Name 2 equipment used for laparoscopic instrument inspection.
-Borescope
-VerifEye
How many air exchanges per hour could be in sterile storage?
10 air exchanges per hour
Exposure time, exposure temperature and dry time for a P2 cycle?
4 minutes, 275 degrees, 30 minutes
When brushing instrument in the sink, where is the brushing supposed to happen?
Under the water level
When should the borescope be used?
During decontamination process.
What 3M stands for?
Minnesota Mining and Manufacturing.
How many air exchanges per hour should Decon have?
Ten air exchanges per hour
What you do when a wrap has visible spotting marks?
Return it for reprocessing.
Which cable is used for two single or dual channeled surgical endoscope or a single channeled endoscope?
Flexible cycle.
What is the maximum fingernail length?
No more than 1/4 inch long.
What information goes in the loaner tags?
-Date of surgery
-surgeons name
-loaner name
Which system should be used to arrange and maintain sterile packages?
First in, First out.
Name of bacterial spores inside of 3M Attest BI?
Bacillus atrophaeus