Which laparoscope size and angle are most commonly used for a Laparoscopic Cholecystectomy?
5mm 30° laparoscope
What are some of the surgical tasks that 3D visualization can assist a surgeon with?
Suturing/knot tying, Dissection
What is the main advantage of modularity in CLICKline instruments?
Maximum hygiene and optimum cleaning
Name one advantage of the AIDA BELLA compared to the STRYKER Connected OR Hub?
Ability to record in 3D, cost, doesn't require license for each functionality, TOMS integration
What is the maximum pressure for high flow mode? What is the max pressure for Pediatric mode?
High Flow: 30 mmHg, Pediatrics: 15 mmHg
Why should you focus on selling Rubina laparoscopes superior to traditional scopes?
Revenue generation, NIR ready, better lens, edge to edge clarity, harder for third party to repair.
What are the key features of the TIPCAM Rubina compared to OLYMPUS 3D Endoeye?
digital horizon control, 4K , ICG
Why might traumatic graspers be used during a case?
Tissue touched needs to be extracted from the body
Which feature allows users to record either one input or both inputs simultaneously in the STORZ software?
Dual Channel
Who are the two biggest competitors in the insufflator market?
Stryker (PneumoSeal) and ConMed (AirSeal)
Name 2 things that could cause damage to a scope during surgery and/or during the sterilization process
Drop, torque, remove from the surgery promptly, not using a sterilization tray, ultrasonic flash, instruments thrown on top
In your own words describe Horizon Control to a surgeon.
the feature maintains the image orientation in an upright position while the TIPCAM is rotated. There is a gravity sensor is used in conjunction with software to maintain image orientation. This only requires a one hand motion unlike a traditional camera and telescope.
What are the two types of materials that are used for the handles of CLICKline instruments?
Plastic and metal
How much internal memory is in the AIDA Bella?
2TB
What are key benefits of including a service contract when selling laparoscopes?
It makes the customer happier and more loyal. It helps your run rate for recurring revenue. It reduces communication issues with the customer
Name 2 procedures that would be good targets for the TIPCAM and why?
Roux-en-Y, hysterectomy, sacrocopopexy, liver resection, why? Because they are already using a large trocar for these types of cases. Also, complexity and it helps with technical tasks such as suturing and dissection- low pelvis, complex anatomy
What is the standard diameter and length for neonate instruments in the CLICKline portfolio?
2mm x 20cm
Name all the ways in which you can trigger images and videos on the AIDA BELLA?
Using the smart screen, from the camera head, and wireless foot pedal
A surgeon is experiencing a blurry or dark image during a laparoscopic procedure. What steps should you take to troubleshoot the issue. Name 2.
Inspect the scope for damage (dents, broken lens) using a loop. Check the light cable for loose connections. Wipe/polish the lens to remove blood or debris. Ensure the light source is set correctly (automatic or at least 50%). Verify monitor settings and adjust as needed. Check if the scope was torqued during the case or if there is a broken light fiber.
You see a double image on the 3D monitor, even with 3D glasses. What’s the most likely cause?
The monitor may be positioned incorrectly. The viewer's line of sight should be perpendicular to the center of the screen. Could also be that the wrong input is being displayed on the monitor, should be display port.
What should you do if there is no print icon on the GUI?
Set the default print setting to 'during capture' or 'Mandatory'