These 2 positions used for the pre-surgical evaluation
Sitting and lying on the left side
100
The potential cardiac conditions that can occur during diagnostic testing in the OR
bradycardia and asystole
100
This diagnostic test must be performed during the procedure to confirm the generator and lead are properly connected.
System Diagnostics
100
The most common reason for high impedance during a NEW implant in the OR
Lead pin not fully inserted into the generator
100
The one item needed during diagnostic testing in surgery NOT provided by LivaNova
a laser arm drape/probe cover/video camera bag
200
This is the minimum R-wave amplitude required for the AspireSR device to sense heart rate.
0.4mV
200
This is the suture color that denotes the positive electrode
white
200
In a typical case, System Diagnostics is performed twice - under these conditions
outside and inside the pocket
200
On a dual pin lead, the positive lead pin is denoted by these two indicators
White marker band and Model/SN
200
This part of the strain relief should be 3cm in circumference and secured parallel to the anchor tether
strain relief bend
300
This machine is used to perform the pre-surgical evaluation
ECG/EKG
300
This is the period of time the device should remain off following implant.
14 days
300
This must be verified in the OR during initial implant of the AspireSR
heartbeat detection
300
This must be used when performing Generator Diagnostics
Test resistor
300
This is the recommended direction for tunneling
Neck to chest
400
This is the placement for the RA electrode
left neck/lead implant position
400
The sterilization indicator should be this shade to indicate the device was sterilized before shipping
gold to bronze
400
High impedance in the OR is generally noted by an impedance value in this range
5300 - 10,000 ohms
400
Failure to communicate with a new generator outside of the sterile field prior to the case likely indicates this
communication/programmer system error
400
This is the position that the patient should be placed in for surgery
supine/lying on the back
500
This is the heartbeat detection sensitivity that corresponds with 0.86-1.25mV amplitude range.
2
500
This surgical equipment should not be used once the generator is introduced in to the sterile field
electrocautery/Bovie
500
This indicates a short circuit or low impedance
Impedance value < 600 ohms
500
Besides lead/pin connector issues, another cause for high impedance in a NEW implant case could be this
electrodes off the nerve
500
The vagus nerve typically is positioned in a posterior groove behind these two vessels
the carotid artery and the jugular vein