XE-Series
XT-Series
XS-Series
SP-1000
HST
100
How many sections are in a XE SRV?
What is 3.
100
Which chamber of the reaction unit measures NRBC on the XT-series instrument?
What is None.
100
How many SRV sections are located in the XS series instrument?
What is None.
100
What causes no methanol being used during daily clean but there are no errors?
What is Methanol line kinked.
100
What causes all racks to pass XE on skip line?
What is Conveyor is in single mode operation.
200
How many chambers are in the XE reaction unit?
What is 4.
200
What is another name for EPK reagent?
What is CellPack.
200
How many chambers are in the XS series reaction unit?
What is 1.
200
During weekly shutdown system gives stain chamber not filled error what reagent is not filling?
What is Methanol.
200
What causes racks to go to the XE but do not spin tubes.
What is Wrong racks being used.
300
What is the third chamber back in the XE reaction unit?
What is DIFF.
300
What is a indicator that the instrument is at it ready state?
What is the Green LED.
300
On a XS analyzer with a auto loader, how many samples can be loaded at once?
What is 20.
300
What causes cassettes to stick together but not caused by overflow?
What is Cassettes not dried after washing.
300
What happens when racks are correct but some tubes don't spin?
What is Labels binding.
400
Where is the .25 pressure adjusted on the XE series analyzers?
What is Pneumatic unit.
400
At how many cycles does the XT-series piercer need to be replaced?
What is 30,000 cycles.
400
What is the leading cause of reaction unit and RBC mixing chamber drain errors?
What is Filter #13 clogged.
400
What causes slides to have 2 stain lines?
What is Clogged pipette at stain 1 or buffer station.
400
Why after the HST line is rebooted the racks sit in start yard?
What is Start switch needs pressed.
500
What waste chamber does all 4 of the XE series reaction unit chambers drain into?
What is Waste chamber 1.
500
What master valve controls waste chamber 2 draining?
What is MV18
500
Which rinse cycle requires the user to aspirate bleach, shutdown sequence or monthly rinse?
What is Monthly rinse.
500
When running in stain only mode, why is there always an empty cassette at end of run?
What is Unit knows when last slide has been processed, this is normal operation.
500
What causes the analyzer to try to sample where no tube exists?
What is Lower sensor not being activated.