PARTS
ALARMS
WATER
INFECTION CONTROL
MISC.
100
WHAT DO THE DIALYZER CONNECTORS DO?
BLUE BRINGS CLEAN DIALYSATE, RED REMOVES DIRTY DIALYSATE TO THE DRAIN.
100
WHAT DOES TMP STAND FOR? 

TRANSMEMBRANE PRESSURE

100
WHAT RATIO OF WATER TO BLEACH DO YOU USE TO CLEAN THE MACHINE?

1 PART BLEACH TO 9 PARTS WATER (1:100)

100

 How long do we wash our hands with soap and water?

40-60 seconds

100

HOW FULL SHOULD THE ARTERIAL CHAMBER BE?

85-95% FULL

200

WHERE IS THE HAND CRANK LOCATED?

BACK OF THE MACHINE

200

WHY MIGHT YOUR MACHINE ALARM TMP?

FLOODED TRANSDUCERS OR CLOTTING SYSTEM.
200
WHEN USING A D6 METER, WHAT RANGE CAN THE CONDUCTIVITY BE BETWEEN TCD AND METER?

+/- 0.3

200

How long can hepatitis B live on surfaces?

Up to 7 days

200

YOU HAVE A FLOODED TRANSDUCER, WHEN YOU CHANGE THEM YOU NOTICE BLOOD WENT THROUGH TO THE OTHER SIDE. WHAT IS YOUR NEXT STEP.

AFTER THAT PATIENTS' TREATMENT, PULL THE MACHINE FOR INTERNAL DISINFECTION.
300

WHAT DEVICE STOPS AIR FROM GETTING TO OUR PATIENTS?

VENOUS OCCLUDER CLAMP

300

WHY IS IT IMPORTANT TO ADDRESS ALARMS IN A TIMELY FASHION?

WHEN THE BLOOD PUMP STOPS, IT INCREASES THE RISK OF CLOTTING IN THE EXTRACORPOREAL CIRCUT

300

WHEN VERIFYING MACHINE CONDUCTIVITY WHAT IS THE ACCEPTABLE RANGE BETWEEN MACHINE AND TCD?

+/- 0.5

300

WHAT PPE IS REQUIRED TO CLEAN A MACHINE?

GOWN, GLOVES, FACESHEILD
300

AIR EMBOLISMS CAN BE AVOIDED BY ENSURING WHAT IS ENGAGED?

OCCLUSION CLAMP (AIR DETECTOR)

400

WHAT DOES THE DIASAFE FILTER DO?

MAKES DIALYSATE ULTRAPURE.

400

WHAT HAPPENS WHEN THERE IS A HIGH PRIORITY ALARM?

RED LIGHT, AUDIBLE ALARM (BEEP, BEEP, BEEP....) BLOOD PUMP STOPS.

400

HOW MANY CHECKS SHOULD BE DONE OF THE MACHINE SETTINGS AND EXTRCORPOREAL SYSTEM WITHIN THE FIRST HOUR OF THE PATIENT'S TREATMENT?

2 PRIOR TO START OF TREATMENT AND (1) RN TO VERIFY WITHIN 1 HOUR OF THE PATIENT STARTING TREATMENT.

400

WHAT IS CONSIDERED A LARGE BLOOD SPILL AND WHAT RATIO OF BLEACH TO WATER DO WE NEED TO CLEAN IT?

>10ML ABSORB, THEN1:10, THEN 1:100

400

TRUE OR FALSE. WHEN RETRIEVING CLEAN SUPPLIES YOU NEED TO BE WEARING CLEAN GLOVES.

FALSE
500

NAME THE PARTS OF THE ECC

NEEDLES OR CVC, DIALYZER, ARTERIAL BLOODLINE, VENOUS BLOOD LINE

500

WHAT COMPLICATION COULD A MACHINE With A HIGH CONDUCTIVITY CAUSE?

HEMOLYSIS, BURSTING OF RED BLOOD CELLS

500

WHAT RANGE MUST THE PH BE BETWEEN?

6.9-7.6

500

In order to prevent contamination, we must spike the saline bag or connect the lines, using ______________________ technique. 

  

ASEPTIC TECHNIQUE

500
WHERE DO YOU GO ON FME-CONNECT TO FIND POLICIES?

POLICYTECH