This Alarm is related to kinks in the Arterial line or poor arterial access flow.
What is the Low Arterial Pressure alarm?
This model Tablo machine has a Minimum Dialysate flow rate of 50ml/min
What is the Tablo PRO+ machine?
This number represents the total amount of Bicarb a patient on Tablo experiences.
What is Total Buffer?
This disinfection cycle is required at the end of each day's use.
What is a heat disinfection?
This Regulatory body provides standards for Water For Hemodialysis And Related Therapies.
What is AAMI (Association for the Advancement of Medical Instrumentation) ?
This Alarm is due to increased pressure as the blood is going back to the patient. This may be caused by kinks or poor venous flow.
What is the High Venous Pressure Alarm?
A dialysate flow rate of 300ml/min represents Tablo's minimum or maximum flow rate
What is Tablo's Max Dialysate flow rate?
This is the range of Sodium that Tablo can deliver.
What is 130-145?
Tablo requires this type of disinfection once every 7 days and it takes 7-9 hours.
What is a Full Chem?
Per Tablo's User Manual, How often does Tablo require Chlorine and Chloramine testing?
What is before every treatment and every 8 hours during treatment?
This alarm happens during blood return when the Big red clamp or saline clamp is engaged
What is the Saline Clamp or Red Clamp Error?
When Dialyzing a large patient using the largest dialyzer, the highest DFR and the highest BFR, this is the only other option to improve clearance.
What is increase time?
This setting allows a nurse to deliver a set amount of saline to the patient on a specified time interval throughout the treatment.
What is a scheduled Saline Flush?
Cleaning the dialyzer connectors and dialysate caps are a part of this type of maintenance.
What is Routine maintenance?
DAILY DOUBLE!!!!
The Empty Bed Contact Time on Tablo is shorter due to a smaller surface area. The EBCT for Tablo is ____
What is 2.08 minutes?
This alarm is triggered when Tablo senses a discoloration in the used dialysate line. It may be elevated bilirubin or blood.
What is Blood in Dialysate alarm?
Tablo has 3 therapy options. IHD, UF Only and ______
What is Sequential Dialysis?
This setting limits how long the blood pump can be stopped before Tablo prohibits blood return.
What is the blood clot timer?
What is the name of the disinfectant that is used to perform Tablo's weekly chemical disinfection process?
Minncare HD
Conductivity and PH testing were a standard for every dialysis treatment. Tablo doesn't require this testing because Tablo has_________
What are 2 redundant Conductivity sensors?
DAILY DOUBLE!!!!!!!
This alarm is caused by a Sensor failure in the top door of the dialysis machine that determines blood in dialysate.
What is a blood leak sensor failure?
A very sick patient in the ICU on multiple pressors may require an extended therapy. What extended therapies can Tablo deliver?
What are SLED and P+ PIRRT or 24 hour therapies?
This device is used to save alarm limits and patient prescription.
What is the Patient key?
This type of disinfection takes 90 minutes and is used between patients that are Hep B+ or Hep status unknown.
What is a Quick Chem?
Tablo's requirements for water are 5-32 degrees Celsius and ______PSI
What is 30-80 PSI water pressure?
This is the Achilles Heel of Dialysis
What is vascular access?