CT/vPad
Spin Planning
Spin Drive
Spin Perc
Instruments
100

How to place pads for lateral scan

vPads will need to be placed on the patient's side that is up or on patients back.  A T or L formation is acceptable 

100

My scan is way to high/low

Auto or manual trim 

100

"Not tracking vpad" site already changed vpads and connector cable 

Field generator is upside down. 

100

My patient was scanned lateral and pads are on the back broke what do I do now?

Pads need to be placed on the side for vpad snapshot

100

Do vpads have an expiration date?

No

200

Site forgot to place pads before scan was done

Either rescan or place pads before procedure and proceed with referral 

200

My tree did not grow out all the way 

Do a manual start (Explain how to do this)

200

When navigating and my instrument is floating outside the airway was can that be an indication of?

Registration is off

200

What is a good amount of Dye if preparing for a Meth Blue dye injection?

.5cc-2.0ccs

200

Spin xtend will not attach lock onto scope

the olympus biopsy valve needs to be attached to the port


Note: which does not come with 

300

CT's would not load onto my planning station from USB/CD or DICOM

CT may have Scout Images imbedded. Resend images or put new images on CD/USB.

Could also be a connection issue



300

I am on my mapping review screen and my red alignment is not over my target 

Do a mapping refinement 

300

My right take off for checking the secondary is slightly off

Lets check the left side--further down the airway 

300

For a 1cm throw with the biopsy gun where am I obtaining the sample?

distal portion 

300

I have a forceps plugged in but software sees it as a needle

Unplug multiple instruments at a time, software is recognizing one verse the other

400

Outward pad arrow was side to side and not cranial caudal

The sensors should be cranial-caudal during the CT. If the pad is placed sideways, then the sensor won't make it into 3 axial slices. Rescan or replace pads before procedure

400

Where do I get screenshots from my planning laptop?

Diagnostic export

400

What (3) items do I need to send vSupport when sending in data?

.Drive (Doctors plan for case)

Diagnostic data (Service>Diagnostic export)

.Zip(Export from Spin Drive tower after case)

sys4000-mp4 file!

400

For a 1cm throw with the FNA needle where am I obtaining the sample?

proximal portion 

400

True or False: Aspiration occurs at the distal tip of the brushes

False 

500

My scanner does not have the Veran Protocol can I do a normal scan 

No, scanner needs to have Veran Protocol to proceed unless a referral scan is being used. 

500

My inspiration scan only has 100 scans and expiration has 624

Resend inspiration

500

Please explain how to use a Double Lumen tube for registration

Start collecting points in the trachea>go down the RMB collect points in the lower airways until green check mark is filled> Stop Collecting> Pull instrument out go down other tube and start collecting points in the LMB lower lobes

500

Can you put a patient in a lateral position for perc after ct was done in supine position?

you can wedge patient slighty but cannot put them in a lateral position after supine ct scan 

500

I can not get my sample out of the needle

Use a stylet, or flush it out