Strips and QC Controls
Glucometer Functionality
Cleaning the Glucometer
Trouble Shooting
Comments
100

Once opened the Test strip vial is good for how many days 

 180 days 

100

Why would you dock the Nova meter 

Charge the battery and to find meter

100

What is used for cleaning the meter

2 bleach wipes 

100

What is the first step if meter not working 

Remove battery for 30 seconds then reinsert

100

Do you have to make a comment in the glucometer if you have a critical value

Yes, you must enter a comment. you can enter up to 3 different comments

200

Once opened the QC vials are good for

90 days

200

Why is the meter kept flat or pointing downward 

To keep the sample from seeping into the test strip insertion slot destroying the meter and no one else can use it

200

What is the dwell time for each bleach wipe

4 minutes

200

If the strip does not fill completely, do you add a second drop

No get a new test strip insufficient sample or sample applied incorrectly

200

What must you do with a critical value

You must order a stat glucose, notify physician

300

How often are QCs run on the Glucometer

At least every 24 hours

300

Which end of the meter has the scanning laser     

Top of the meter

300

Wiping the external surface of the meter thoroughly is which step in the 2-step process

Cleaning step

300

Do you send the meter through the tube system if not working

No never send through the tube system walk the meter to the lab

300

How do you know if you have a critical value on the meter

Red double arrows up or down

400

Why should you keep the test strip bottle closed and listen for a snap

Strips affected by light, heat, humidity

400

Which end of the meter holds the test strip

Bottom left of the meter

400

what step is thoroughly wiping the surface of the meter top, bottom, left, right, test strip port, minimum of 3 times vertically avoiding the bar code scanner and electrical connector

Disinfecting step

400

What does the error message transfer failed mean on the meter

meter was undocked prior to data transfer must redock prior to using meter

400

If you see HI or Low instead of a number what does that mean

Hi if value is greater than 600 and Low if value is less than 10

500

Where would you find new test strip vials and QC vials when empty

Lab   

500

Where is the battery for the meter

on the back of the meter

500

How often is the meter cleaned

After each patient use

500

How do you keep your meter access current

 Complete HS and posttest and perform QC on the meter. Or contact the lab point of care coordinator 

500

What comment should you use in the meter after every patient use

"Cleaned" meter comment after meter has been cleaned and disinfected with 2 bleach wipes with a dwell time of 4 minutes each