Cleaning the machine externally requires this solution using a damp cloth
What is 1:100 bleach solution?
Another name for Hansens
What is dialyzer Connectors?
Choose this Menu Option to deliver a bolus or change tx Modality.
What is the Online Menu?
This component delivers Substitution fluid
What is the Substitution Fluid Port?
Place to check for residual bleach
What is at the drain?
Citric Acid is Being replaced with this premixed solution
What is citrasure?
These Doors are removable and easily reattached.
What are EBM Doors?
Use this menu option to stop the blood pump to exit priming prior to connecting patient.
What is Preparation Menu?
This replaces Normal Saline (NS) on 5008x Machine
What is Substitution Fluid?
If using a BiBag this test is no longer required for 5008X machine
This is completed every 7 days
What is cold (bleach) disinfection?
Requires an exchange every 90 days OR after 100 heat disinfect OR after 13 cold disinfect
What are the Diasafe Filters
Choose this Menu Option for viewing the instructions for machine set-up.
What is Blood System option?
The amount of substitution fluid it takes for priming
What is 500ml?
3 Components of Dialysate
What is Acid, Bicarb and RO (treated) water?
This is completed daily at end-of-shift
What is heat disinfect?
This takes the place of Arterial Blood Chamber
What is the Arterial Pressure Measurement Unit?
Buttons that are selectable and allows changes to be made
What is blue?
This creates a sterile and non-pyrogenic substitution fluid.
What is the function of diasafe filter #2?
This is an independent test performed by the 5008x machine testing for electrical currents of dialysate.
What is Conductivity?
This is required when contamination is suspected with substitution port or rinse port.
What is a mandatory Disinfect?
Must be used with 60cc syringe to open artertial pressure measurement unit after power outage
What is the Check Valve?
Color of Bar during T1 Test, Preparation and Prime
What is orange?
Name 5 Components of Online Plus System
What is: Substitution Fluid Pump, Rinse Port, Substitution port, Diasafe 1&2
Indicates Dialysate is flowing?
What are the green waves with actively moving gray bar?