Parts
Menus
Online
Water
Disinfection
100

The new name for "Hansens"

What is Dialyzer Connectors

100

Color of buttons that are selectable, allowing changes to be made

What is blue

100

This replaces normal saline on the 5008x machine

What is Substitution Fluid

100

Location to check for residual bleach

What is wall drain

100

External cleaning of the machine requires this solution using a damp (NOT saturated) cloth 

What is 1:100 bleach solution 

200

This takes place of the Arterial Blood Chamber

What is the Arterial Pressure Dome

200

Choose this menu for viewing the machine set-up instructions

What is Blood System 

200

The amount of substitution fluid the machine will prime with 

What is 500mL 

200

If using a bibag, this test is no longer required for the 5008x

What is pH
200

This is completed every 7 days 

What is cold disinfect (bleach)

300

These doors are easily removable and replaced 

What is EBM doors (extracorporeal blood circuit module doors)

300

Go to this menu option prior to connecting to patient, in order to stop blood pump and exit priming

What is Preparation 

300

This piece provides the substitution fluid

What is Substitution Fluid Port

300

Name the 3 components of dialysate

RO, acid, bicarb

300

This is completed at the end of every day

What is chemical heat disinfect

400

These require an exchange every 90 days, 100 heat cycles, OR 13 cold disinfects (bleach) - whichever comes first

What is the Diasafe Filters

400

OCM, AFM, BPM, Circulation - are all found under which menu

What is Options Menu

400

This creates a sterile, non-pyrogenic substitution fluid 

What is Diasafe Filter #2

400

This is an independent test performed by the 5008x, that tests for electrical currents of dialysate

What is conductivity

400

This is required when contamination of either port is suspected

What is 51 minute mandatory disinfection 

500

This must be used with 60cc syringe to open the Arterial Pressure Measurement Unit after a power outage 

What is the Check Valve

500

When delivering a substitution fluid bolus or changing tx modality, we would choose this menu 

What is Online

500

List the 6 components that make up the Online System

What is Diasafe Filter #2, Substitution Fluid, Substitution Fluid Pump, Substitution Fluid Port, Rinse Port, Safeline

500

This indicates that dialysate is on and flowing 

What is the green waves with white moving bar

500

Citric Acid is being replaced with this premixed solution

What is CitraSure