This Stakeholder is most important in the implementation of Expanded Access.
Who is the Surgeon Champion?
This is how Intuitive's turnover time definition typically differs from the Hospital's
What is da Vinci undock to dock vs Patient Wheels Out to Wheels in Turnover
This is the minimum amount of crucial inventory to keep on hand.
What is PAR?
These are the electrical requirements for an Xi system
what are3 dedicated circuits (4 if dual consoles) with 15amps each?
What are the steam sterilization parameters for the Ion catheter guide?
What is 270F, 4 minutes exposure, and 20-30 minutes dry time.
This is the typical block release time for benign case types such as URO and GYN
What is two weeks?
Genesis recommends these parallel tasking activities with the patient in the room.
What is drape da Vinci?
What is setup the backtable?
What is count?
What is five more than the OR lives?
These are the low temp sterilizers validated for Xi endoscopes
These are the hangtime requirements for Ion products
What is it is up to hospital policy, Intuitive does not have a recommendation
These are the reasons why Genesis da Vinci utilization calculations are usually lower than that of hospital utilization calculations?
What is da Vinci utilization does not include 1st case setup to docking time?
Genesis recommends the room is ready for the patient when
Equipment checked.
Anesthesia ready.
Bed ready.
Backtable pack opened and tech is scrubbed in.
Trays, Instruments, Disposables available.
When calculating par levels, what is the mode and how do you select it?
The mode is the most number of cases performed in one day that is not a statistical outlier, and is found in the carrying cost tab of excel.
These Five AER's are Validated for Ion Reprocessing
What are Medivator Advantage Plus, Medivator Advantage Pass Through, Medivator DSD Edge, Steris 1E, Steris 1 Endo
What is the max temperature allowed for the Xi endoscope?
What is 140F.
This is the maximum recommended console time and the minimum recommended number of cases needed for flip rooms.
What is less than 90 min and at least 5 cases?
Here are three examples of verbal/electronic communication best practices Genesis recommends for the end of the case.
Surgeon communicating with Anesthesia approximate case end time.
RN communicating with EVS approximate wheels out time.
Surgeon communicating with OR Staff his/her target return to room time.
OR Staff communicating with SPD on tray, instrument, or endoscope rapid turnaround requests.
RN communicating with Pre-op next patient status.
When calculating par levels, what is the standard selection for turnover time?
What is 24 hours.
The SP system will fit in most elevators, as long as they are this wide
What is 41in wide?
This is the most common reason for disc breakage
What is an enzymatic issue?
These are three ways to maximize da Vinci access on a daily basis
What is stack da Vinci cases during their all-day robotic block and stack non-da Vinci cases during other block times.
What is front load robotic cases and putting other procedures at end of the day.
What is movie lap procedures that can be done with da Vinci to da Vinci room during the day when da Vinci becomes available.
What is schedule the same procedure type and room configurations back-to-back, reducing workflow inefficiencies time and the need to rearrange a room multiple times a day.
These are four examples of surfaces/equipment that are "enhanced" or "terminal" cleaned, but do not need to be cleaned during each turnover.
What are:
steps, workstation on wheels, garbage or linen bins, stools, walls, doors.
For an instrument with 15 lives, what is the minimum utilization percentage to be recommended for trays.
What is 75% (15 OR lives divided by 20 reprocessing lives).
This is the process ultrasonic's use to clean instrumentation, and how it works.
What is Cavitation, and by sending vibrations through the water, which create bubbles that implode, breaking up debrief?
This Amount of Air pressure should be applied when doing a leak test for the SP Camera
What is 140-180mmHg (19-24kpa)?