100

The default gas for FC200

Oxygen

100

Name all the components included on the Cellicon tubing assembly. 

Aseptiquik connectors, Levitronix feed pump, 3X pressure sensors, SU permeate flowmeter. 

100

EM stands for...

Equipment Module

100

To reduce or eliminate oscillations of the present value about the setpoint, I would adjust this parameter. 

I value - Reset

100

The highest cell density that has been achieved in the iFlex system, so far.

126e6 cells/mL

200

The preconfigured function for Pump P007 on the perfusion tower. 

Cell Bleed

200

The number of small addition lines provided on top of the bag. 

5

200

List the control modes, in order of increasing automation.

MAN/ROUT, AUTO, CAS/RCAS

200

The minimum working volumes of the 200L and 2000L bioreactors. 

40L and 400L.

200

The approximate kLa of each sparger. 

Open pipe = 5

Mid-range = 50

High performance = 100

300

The number of E-Stops available on the system

3

300

This sparger is located on the left-hand side of the 200L bag, when placed into the tank. 

Mid-range / drilled hole

300

For certain PID control loops, this mode indicates that the PID algorithm is not used by the module. Instead, the setpoint is used to immediately determine an output.

bypass mode

300
Name 3 parameters that can be used to monitor performance of the cell retention device. 

TMP, Pressure Drop, Flux, Throughput

300

All bioreactors achieve less than this mixing time. 

34 seconds

400

The MFC output options for the optional FC500, FC600, and FC700 on the 200L system. 

5, 10, 20 or 50 SLPM

400

Two potential issues during the bag installation process. 

Many will be accepted...

400

Name 3 purposes for the engineering laptop. 

User management, recipe development, troubleshooting

400

Name 7 of the available calibration workflows. 

pH, DO, CO2, VCD, Flow sensor zero, pressure sensor zero, pump flow control, pump speed control, tare weight 

400

The location of the highest shear zones within a stirred tank bioreactor. 

The impeller zone

500

The accuracy of the DO sensor, pH sensor, and Temperature probe

DO +/- 10% of MV; pH +/- 0.1; Temp +/-0.2 degC

500

Name each of the available ports on the side panels of the bag. 

Left: 2X DO slots, 2X pH slots, Temp slot, 2x sampling lines

Right: VCD slot, 3x MU probe slots, cell bleed, CRD return, 2X sampling lines

500

Explain the difference between a phase, operation, and unit procedure. 

A Unit Procedure defines the strategy for carrying out a contiguous process within a unit. (Bioreactor Perfusion Process) An Operation contains one independent processing activity. (Liquid Addition) A Phase is the specific occurrence of an operation. (Antifoam Addition)

500

List the steps of executing an EM for DO Operation. 

1.Select SP, Flex Profile, Device Assignments, and Gas Destinations from Operating Parameters. 2.Run the Enable Command from DO EM. 3. Verify DO EM Command Completes

500

Two sources of deviation in the control performance, experienced during internal cell culture runs. 

Antifoam addition - to DO in fed-batch run; Feed additions using cold media - to temp in fed-batch run; PLC issue - perfusion run